Lab2000
Box body
* Dimensions: 1220mm×750mm×900mm (length×width×height)
* Material: stainless steel structure (Type 304), thickness 3mm
* Filter: aperture 3 microns
* Gloves: butyl gloves with thickness 0.4mm (mouth diameter of 203mm), hard aluminum alloy glove mouth (diameter of 220mm, after corrosion prevention treatment, anodized aluminum)
* Window: tilted operating surface, detachable 8mm safety tempered glass front window, wear resistance, corrosion resistance, good light transmission, OMEGA sealing ring with 3/8 inch thick
* Can maintain a certain positive and negative pressure in usage (-10mbar-10mbar)
* Equipped with anti-reflection lighting energy-saving lamp, the light is soft
* Provided with a double-layer display board
* Provided with a porous power wiring board (power supply: 220V±10% 50Hz±10%)
* Configured with 3 standard KF-40 interfaces
* Leakage rate: ≤0.05vol%/h
Transition chamber
* Shape: cylindrical (304 stainless steel), connected to glove box, controlled by valve
* Diameter 360mm, length 600mm large transition chamber, piston door, sliding tray in transition chamber, polishing, automatic valve controls air pumping and replenishing. A small transition chamber with a diameter of 150mm and a length of 300mm is provided in the box body, manual valve controls air pumping and replenishing.
* Tray: stainless steel 300mm×450mm, can move and extend freely (equipped in the large transition chamber)
* Vacuum degree ≤1bar(vacuum display is required for both large and small transition chamber)
Control system
* Mode: including self - diagnosis, power off - start feature, with pressure control and self - adaptive function. Automatic control, cycle control, password protection, vacuum chamber control adopt LCD display, unit control adopts PLC touch screen.
* Pressure control: control the pressure of the box and the transition chamber at a certain set value
* Pedal: can control the pressure of glove box, easy operation
Circulation system
* Control: water and oxygen are controlled at less than 1PPm
* Adjusting device: PLC touch type adjustment (heating, evacuation, purification, regeneration), the whole process without human monitoring except for 2.5 hours of ventilation
* Working gas: N2 or Ar gas (gas cycle)
* Circulating fan: 0 - 100m3 high speed frequency conversion fan, microprocessor control, pressure tested stainless steel device
* Purification column: single purification column, imported purification materials
* Purification valve: automatic electro-pneumatic valve
* Filter: High efficiency filter with 0.3 micron aperture sieve conforming to HEPA standard
* Regeneration: used for purification system reuse to make it active; regeneration requires a mixture of N2/H2 or Ar/H2, with H2 accounting for 10%
Display system
Unit control adopts PLC touch screen
PLC oxygen analyzer
* Range: 0-1000 PPm
* Error: ±1% PPm
* Environment: -10℃-50℃
PLC water analyzer
* Range: 0-1000 PPm
* Error: ±1% PPm
* Environment: -10℃-50℃
Organic solvent adsorption system
Diameter 250mm*500mm, with vacuum, built-in high quality activated carbon
One side simplex station standard glove box Lab2000
Series No Configuration Qty (Set) 1 Box body 1 2 Transition chamber 1 3 Control system 1 4 Circulation system (including purification and regeneration system) 1 5 Display system 1 6 Gloves 1 7 Pipeline interface 1 8 Press gauge 1 9 PLC water analyzer 1 10 PLC oxygen analyzer 1 11 Organic solvent adsorption system 1 12 Vacuum pump 1